FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 3984552
·
Received August 6, 2014
Report
- Report Number
- 3004209178-2014-90528
- Event Type
- Malfunction
- Date Received
- August 6, 2014
- Date of Event
- June 27, 2014
- Report Date
- July 6, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Additional Manufacturer Narrative · 1
THE INSULIN PUMP WAS RECEIVED WITH BLANK DISPLAY DUE TO MOISTURE DAMAGE ON ELECTRONIC ASSEMBLY. THE INSULIN PUMP WAS UNABLE TO VERIFY BUTTON KEYPAD ANOMALY DUE TO BLANK DISPLAY. THE INSULIN PUMP HAD A CRACKED DISPLAY WINDOW CORNER AND CRACKED RESERVOIR TUBE LIP.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT THERE IS FLUID UNDER THE LCD DISPLAY. CUSTOMER STATES THAT DUE TO THE INSULIN PUMP BEING EXPOSED TO FLUID, THE KEYPAD IS HAVING ISSUES. THE BLOOD GLUCOSE READING IS 104 MG/DL. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 459931 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 23 YR |