FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3984545 · Received August 6, 2014

Report

Report Number
3004209178-2014-90358
Event Type
Malfunction
Date Received
August 6, 2014
Date of Event
July 6, 2014
Report Date
July 6, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP PASSED DISPLACEMENT TEST AND SELF TEST. NO UNEXPECTED ERROR ALARM OR BLACK SCREEN NOTED. THE DEVICE HAD MINOR SCRATCHED LCD WINDOW AND CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED BY THE CUSTOMER THAT THEIR INSULIN PUMP'S DISPLAY SCREEN IS BLANK. CUSTOMER STATES THAT THEY REMOVED THE BATTERIES AND REINSERTED TWO HOURS LATER BUT THE DISPLAY DID NOT RETURN. CUSTOMER STATED THEY DO NOT RECALL ANY SIGNIFICANT EVENTS THAT LEAD TO THE ALARM OR IF THE DEVICE WAS DROPPED. ADVISED CUSTOMER TO DISCONTINUE USE OF DEVICE AND REVERT TO BACK UP PLAN. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 66 MG/DL AT TIME OF REPORTING. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
459597 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523LNAS

Patients

Seq Age Sex Outcome Treatment
1 28 YR