PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-90358
- Event Type
- Malfunction
- Date Received
- August 6, 2014
- Date of Event
- July 6, 2014
- Report Date
- July 6, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP PASSED DISPLACEMENT TEST AND SELF TEST. NO UNEXPECTED ERROR ALARM OR BLACK SCREEN NOTED. THE DEVICE HAD MINOR SCRATCHED LCD WINDOW AND CRACKED RESERVOIR TUBE LIP.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
IT WAS REPORTED BY THE CUSTOMER THAT THEIR INSULIN PUMP'S DISPLAY SCREEN IS BLANK. CUSTOMER STATES THAT THEY REMOVED THE BATTERIES AND REINSERTED TWO HOURS LATER BUT THE DISPLAY DID NOT RETURN. CUSTOMER STATED THEY DO NOT RECALL ANY SIGNIFICANT EVENTS THAT LEAD TO THE ALARM OR IF THE DEVICE WAS DROPPED. ADVISED CUSTOMER TO DISCONTINUE USE OF DEVICE AND REVERT TO BACK UP PLAN. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 66 MG/DL AT TIME OF REPORTING. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 459597 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-523LNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR |