FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3984537 · Received August 6, 2014

Report

Report Number
3004209178-2014-90529
Event Type
Malfunction
Date Received
August 6, 2014
Date of Event
July 6, 2014
Report Date
July 6, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

INSULIN PUMP HAD AN INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACE. NO BUTTON ERROR ALARM MINOR SCRATCHES ON LCD WINDOW, CRACKED DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP AND MISSING END CAP STICKER.

Description of Event or Problem · 1

CUSTOMER REPORTED BUTTON ERROR ALARM. CUSTOMER STATES THAT SHE WAS ABLE TO CLEAR THE ALARM IN THE PAST BY REMOVING THE BATTERY BUT NOT THIS TIME. THE BLOOD GLUCOSE READING IS 487 MG/DL. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
459927 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAH

Patients

Seq Age Sex Outcome Treatment
1 25 YR