PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-90348
- Event Type
- Malfunction
- Date Received
- August 6, 2014
- Date of Event
- July 6, 2014
- Report Date
- July 6, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY
- Reporter Occupation
- PATIENT
Narratives
THE INSULIN PUMP ALARMED FOR A BUTTON ERROR AND HAD NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. THE INSULIN PUMP HAD NO MOISTURE DAMAGE ON THE ELECTRONICS, MOTOR, VIBRATOR MOTOR AND BATTERY TUBE ASSEMBLY. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON THE DISPLAY WINDOW, SCRATCHED RESERVOIR TUBE WINDOW AND STAINED END CAP STICKER.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE CUSTOMER'S BLOOD GLUCOSE WAS UNKNOWN. THE CUSTOMER STATED THAT THE DEVICE WAS ON THEIR BRA AND THEY WERE SWEATING. THEY TRIED REPLACING THE BATTERY BUT THE ISSUE PERSISTED. THE DISCONTINUATION OF THE DEVICE WAS ADVISED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 459538 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR |