FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3984503 · Received August 6, 2014

Report

Report Number
3004209178-2014-90351
Event Type
Malfunction
Date Received
August 6, 2014
Date of Event
July 6, 2014
Report Date
July 6, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP PASSED THE REWIND, BASIC OCCLUSION, OCCLUSION, PRIME, EXCESSIVE NO DELIVERY ALARM AND DISPLACEMENT TEST. NO UNEXPECTED EXCESSIVE NO DELIVERY ALARM. NO MOTOR ERROR ALARM NOTED. THE MOTOR PASSED THE MOTOR TEST. MINOR SCRATCHED LCD WINDOW, CRACKED CASE NEAR THE DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBE THREADS AND CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. THE ALARM WAS CLEARED BUT THEN THE DEVICE ALARMED NO DELIVERY, TWICE. THE CALLER TRIED PRIMING THE INSULIN PUMP AND THE DEVICE ALARMED MOTOR ERROR AGAIN. THE CALLER STATED THAT THE CUSTOMER WAS PLAYING IN DIRT AND SAND. THE CUSTOMER DOES NOT USE THE SENSOR FEATURE. THE CALLER WAS ABLE TO REWIND THE INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE WAS 568 MG/DL, WHICH WAS NOT TREATED. THE DISCONTINUATION OF THE DEVICE WAS ADVISED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
459531 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAL

Patients

Seq Age Sex Outcome Treatment
1 4 YR