FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3984437 · Received August 6, 2014

Report

Report Number
3004209178-2014-89985
Event Type
Malfunction
Date Received
August 6, 2014
Date of Event
July 7, 2014
Report Date
July 7, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Report Source
Manufacturer report
Reporter Location
MA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS STATES THAT THE INSULIN PUMP RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM DURING TESTING. ALSO RECEIVED WITH CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, SCRATCHED LCD WINDOW AND STAINED END CAP STICKER.

Description of Event or Problem · 1

CUSTOMER REPORTS A BUTTON ERROR ALARM ON THE INSULIN PUMP, WHICH OCCURRED WHILE TRYING TO DELIVER BOLUS. CUSTOMER STATES AFTER ALARM KEYPAD BECAME UNRESPONSIVE. CURRENT BLOOD GLUCOSE LEVEL IS 148 MG/DL. ADVISED CUSTOMER INSULIN PUMP WILL BE REPLACED AND REVERT TO A BACK-UP PLAN PER HEALTH CARE PROVIDER'S INSTRUCTION. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
461190 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAP

Patients

Seq Age Sex Outcome Treatment
1 18 YR