FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3984427 · Received August 6, 2014

Report

Report Number
2531779-2014-22586
Event Type
Malfunction
Date Received
August 6, 2014
Report Date
July 26, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 10/06/2014. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE PUMP POWERS WITH RETURNED BATTERY CAP TO A DIM DISCOLORED DISPLAY. BATTERY CAP IS ABLE TO FULLY TIGHTEN. BATTERY COMPARTMENT CRACK OBSERVED IN THREAD AREA. THERE WAS EVIDENCE OF MOISTURE CONTAMINATION INSIDE BATTERY COMPARTMENT. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO REBOOT, LOSS OF POWER OR CALL SERVICE ALARMS DUPLICATED. THE KEYPAD WAS INTACT. ALL KEYS WERE INTERMITTENTLY UNRESPONSIVE. EVALUATION REVEALED THAT THERE WAS CONTAMINATION UNDER ALL KEY CONTACTS. LEAK TEST SHOWS LEAK AT BATTERY COMPARTMENT CRACK. REMOVED PUMP CASE, EVIDENCE OF ADDITIONAL MOISTURE CONTAMINATION INSIDE PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING MOISTURE INGRESS, POWER ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE DEVICE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
460844 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 31 YR