PARADIGM INSULIN INFUSION PUMP
MDR report key: 3984418
·
Received August 6, 2014
Report
- Report Number
- 3004209178-2014-89983
- Event Type
- Malfunction
- Date Received
- August 6, 2014
- Date of Event
- July 7, 2014
- Report Date
- July 7, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- K040676
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TU
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS STATES THAT THE INSULIN PUMP PASSED THE DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME/A33 AND EXCESSIVE NO DELIVERY TEST. NO MOTOR ERROR ALARM NOTED. MOTOR PASSES MOTOR TEST. ALL BUTTON FUNCTIONING PROPERLY. NO BUTTON ERROR ALARM NOTED. HOWEVER, FOUND CORRODED KEYPAD TRACES. PUMP RECEIVED WITH CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, SCRATCHED LCD WINDOW, SCRATCHED RESERVOIR TUBE WINDOW AND MISSING END CAP STICKER.
Description of Event or Problem · 1
IT IS REPORTED THAT THE CUSTOMER RECEIVED A MOTOR ERRORS ON HIS INSULIN PUMP. CUSTOMER CURRENT BLOOD GLUCOSE LEVEL WAS NOT PROVIDED. INSULIN PUMP WILL BE REPLACED FOR THE CUSTOMER. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 460842 | PARADIGM INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-715WWL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |