FDA Adverse Event Malfunction Summary report: N

PARADIGM INSULIN INFUSION PUMP

MDR report key: 3984418 · Received August 6, 2014

Report

Report Number
3004209178-2014-89983
Event Type
Malfunction
Date Received
August 6, 2014
Date of Event
July 7, 2014
Report Date
July 7, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
K040676
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS STATES THAT THE INSULIN PUMP PASSED THE DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME/A33 AND EXCESSIVE NO DELIVERY TEST. NO MOTOR ERROR ALARM NOTED. MOTOR PASSES MOTOR TEST. ALL BUTTON FUNCTIONING PROPERLY. NO BUTTON ERROR ALARM NOTED. HOWEVER, FOUND CORRODED KEYPAD TRACES. PUMP RECEIVED WITH CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, SCRATCHED LCD WINDOW, SCRATCHED RESERVOIR TUBE WINDOW AND MISSING END CAP STICKER.

Description of Event or Problem · 1

IT IS REPORTED THAT THE CUSTOMER RECEIVED A MOTOR ERRORS ON HIS INSULIN PUMP. CUSTOMER CURRENT BLOOD GLUCOSE LEVEL WAS NOT PROVIDED. INSULIN PUMP WILL BE REPLACED FOR THE CUSTOMER. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
460842 PARADIGM INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-715WWL

Patients

Seq Age Sex Outcome Treatment
1