FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 3984415
·
Received August 6, 2014
Report
- Report Number
- 3004209178-2014-89979
- Event Type
- Malfunction
- Date Received
- August 6, 2014
- Date of Event
- July 7, 2014
- Report Date
- July 7, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS STATES THAT INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE AND BUTTON ERROR. ALARM WAS DUE TO MOISTURE DAMAGE ON KEYPAD TRACES. NO FROZEN DISPLAY NOTED. PUMP WAS RECEIVED WITH A SCRATCHED LCD WINDOW, CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP AND MISSING END CAP STICKER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE CUSTOMER'S BLOOD GLUCOSE WAS 300 MG/DL. THE DISCONTINUATION OF THE DEVICE WAS ADVISED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 460768 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |