FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3984375 · Received August 6, 2014

Report

Report Number
3004209178-2014-90516
Event Type
Malfunction
Date Received
August 6, 2014
Date of Event
July 6, 2014
Report Date
July 6, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP PASSED THE REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME TEST, EXCESSIVE NO DELIVERY TEST AND DISPLACEMENT TEST. NO MOTOR ERROR ALARMS NOTED. HOWEVER, THE MOTOR HOME SWITCH WAS FOUND CORRODED DURING VISUAL INSPECTION. THE DEVICE WAS RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNERS. THE UNIT WAS RECEIVED WITH SEVERELY SCRATCHED LCD WINDOW.

Description of Event or Problem · 1

CUSTOMER REPORTED A MOTOR ERROR ALARM ON THE INSULIN PUMP AND STATED THAT THEY ARE UNABLE TO REWIND. CUSTOMER'S CURRENT BLOOD GLUCOSE READINGS ARE 143 MG/DL. NO FURTHER INFORMATION REPORTED.