FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3984354 · Received August 6, 2014

Report

Report Number
2032227-2014-05991
Event Type
Malfunction
Date Received
August 6, 2014
Date of Event
July 4, 2014
Report Date
July 4, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. INSULIN PUMP RECEIVED WITH CRACKED CASE DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS AND CRACKED RESERVOIR TUBE LIP. INSULIN PUMP RECEIVED WITH UPDATED SOFTWARE CONFIGURATION.

Description of Event or Problem · 1

IT IS REPORTED THAT THE BUTTONS ON THE INSULIN PUMP ARE NOT RESPONDING. THE PATIENT'S BLOOD GLUCOSE LEVEL WAS NOT RECORDED. PUMP HAS BEEN REPLACED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
461848 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-554LWWP