PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3984306
·
Received August 6, 2014
Report
- Report Number
- 2032227-2014-06199
- Event Type
- Malfunction
- Date Received
- August 6, 2014
- Date of Event
- September 2, 2011
- Report Date
- July 7, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
INSULIN RECEIVED WITH INTERMITTENT BUTTONS DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS NOTED. CRACKED BATTERY TUBE THREADS AND SCRATCHED LCD WINDOW NOTED.
Additional Manufacturer Narrative · 1
THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS WERE NOTED. THE INSULIN PUMP WAS RECEIVED WITH CRACKED BATTERY TUBE THREADS AND A SCRATCHED LCD WINDOW.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A BUTTON ERROR ALARM ON THE INSULIN PUMP. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE PUMP AND REVERT TO A BACK-UP PLAN. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 461322 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-754LWWP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |