FDA Adverse Event Injury Summary report: N

TRIDENT PSL HA CLUSTER 58MM

MDR report key: 3984276 · Received August 6, 2014

Report

Report Number
0002249697-2014-03030
Event Type
Injury
Date Received
August 6, 2014
Date of Event
July 18, 2014
Report Date
July 18, 2014
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
MEH
PMA / PMN Number
K983382
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PATIENT IS (B)(6) INCHES IN HEIGHT. AN EVENT REGARDING REVISION DUE TO A CYST INVOLVING A TRIDENT PSL HA CLUSTER 58MM WAS REPORTED. THE EVENT WAS NOT CONFIRMED. DEVICE HISTORY REVIEW: INDICATED THAT THE SPECIFIED LOT WAS ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. COMPLAINT HISTORY REVIEW: INDICATED THAT THERE HAVE NOT BEEN ANY OTHER EVENTS FOR THE SPECIFIED LOT. CONCLUSIONS: THE REPORTED EVENT INVOLVES THE REMOVAL OF THE PATIENT¿S SHELL DUE TO A CYST BEHIND THE SHELL. THE ROOT CAUSE OF THIS EVENT IS THE CYSTIC LESION IN THE ACETABULUM. THIS FACTOR IS PATIENT RELATED AND NOT DEVICE RELATED.

Additional Manufacturer Narrative · 1

AN EVALUATION OF THE DEVICE CANNOT BE PERFORMED AS THE DEVICE WAS RETAINED BY THE SURGEON AND WAS NOT RETURNED TO THE MANUFACTURER. ADDITIONAL INFORMATION HAS BEEN REQUESTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE REPORTED IN A SUPPLEMENTAL REPORT UPON COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT DEVELOPED A CYST BEHIND HIS ACETABULAR SHELL. DOCTOR WAS ABLE TO REMOVE HIS ACETABULAR SHELL AND REAM TO REMOVE THE AFFECTED AREA. HE THEN REIMPLANTED A TRITANIUM SHELL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT DEVELOPED A CYST BEHIND HIS ACETABULAR SHELL. DOCTOR WAS ABLE TO REMOVE HIS ACETABULAR SHELL AND REAM TO REMOVE THE AFFECTED AREA. HE THEN REIMPLANTED A TRITANIUM SHELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
461291 TRIDENT PSL HA CLUSTER 58MM IMPLANT MEH STRYKER ORTHOPAEDICS-MAHWAH 25165104

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention