FDA Adverse Event Malfunction Summary report: N

3MM DIAMONHD BALL, 10CM

MDR report key: 3984217 · Received December 18, 2013

Report

Report Number
1045834-2013-11817
Event Type
Malfunction
Date Received
December 18, 2013
Date of Event
August 4, 2011
Report Date
August 4, 2011
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBE
PMA / PMN Number
K042783
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

RELIABILITY ENGINEERING EVALUATED THE DEVICE, AND IT WAS DETERMINED THAT THE ATTACHMENT HAD A WORN/DAMAGED BEARING IN THE DISTAL END. THE DAMAGED BEARING ALLOWED THE CUTTER SHAFT TO MAKE CONTACT THE MIDDLE SUB-ASSEMBLY OF THE ATTACHMENT, WHICH CAUSED IT TO SHEAR. BEARING DAMAGE IS OFTEN THE RESULT OF NORMAL WEAR OUT OVER TIME, BUT CAN BE EXACERBATED BY IMPROPER OR INEFFECTIVE CLEANING AND MAINTENANCE. (B)(4).

Description of Event or Problem · 1

REPORTER FROM THE U.S REPORTED THAT THE CUTTER BROKE DURING SURGERY AND WAS DIFFICULT TO REMOVE FROM THE DEVICE. THE REPRESENTATIVE WAS UNABLE TO RETRIEVE BEARING SLEEVE. THERE WERE NO REPORTS OF INJURY, HOWEVER, IT IS UNK IF MEDICAL INTERVENTION HAD OCCURRED. THERE WAS NO ADD'L INFO AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
663850 3MM DIAMONHD BALL, 10CM HBE DEPUY SYNTHES POWER TOOLS E063045575

Patients

Seq Age Sex Outcome Treatment
1 NI