FDA Adverse Event Malfunction Summary report: N

SUPER SUCKER SHORT CURVED (BX/20)

MDR report key: 3984073 · Received December 18, 2013

Report

Report Number
1045834-2013-16215
Event Type
Malfunction
Date Received
December 18, 2013
Date of Event
August 10, 2012
Report Date
August 20, 2012
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
KDC
PMA / PMN Number
K822065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED BY DEPUY SYNTHES POWER TOOLS. THE DEVICE WAS EVALUATED AND THE CUSTOMER'S COMPLAINT OF PACKAGING DEFECTS COULD NOT BE CONFIRMED. VISUAL INSPECTION ACCEPTANCE CRITERIA ARE "NO LOOSE FOREIGN MATERIAL (LFM) WHEN INSPECTED AT 1X MAGNIFICATION." NO LFM COULD BE FOUND ON DEVICE WHEN INSPECTED AT 10X MAGNIFICATION. ALSO, THE PACKAGE IS IN GOOD CONDITION WITH NO DEFECTS OF THE PEELABLE OR TERMINAL SEAL. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SENT. REF: (B)(4).

Description of Event or Problem · 1

REPORT 3 OF 48. REPORT RECEIVED FROM (B)(6) STATING THAT THE DEVICE HAD DEBRIS IN STERILE PACKAGE. THE DEVICE WAS NOT BEING USED DURING SURGERY. NO INJURY OR MEDICAL INTERVENTION OCCURRED. NO ADDITIONAL INFORMATION PROVIDED. THE DATE OF EVENT WAS UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
663081 SUPER SUCKER SHORT CURVED (BX/20) KDC DEPUY SYNTHES POWER TOOLS F133063259

Patients

Seq Age Sex Outcome Treatment
1