G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2014-23754
- Event Type
- Malfunction
- Date Received
- August 6, 2014
- Date of Event
- July 11, 2014
- Report Date
- July 11, 2014
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
(B)(4). SUBSEQUENT TO THE INITIAL MDR, THE COMPLAINT TRANSMITTER DEVICE WAS NOT RETURNED TO DEXCOM. THE RECEIVER DEVICE (STR-GL-001/(B)(4)), USED WITH THE COMPLAINT TRANSMITTER DEVICE, WAS RETURNED ON (B)(4) 2014. THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION. REVIEW OF THE DATA LOG CONFIRMED THE REPORTED PERMANENT LOSS OF ANTENNA. HOWEVER, THE REPORTED FAULT COULD NOT BE REPRODUCED WITH RECEIVER.
(B)(4).
PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 TO REPORT AN OUT OF RANGE SIGNAL ON (B)(6) 2014. PATIENT WAS USING A TRANSMITTER THAT HAD EXPIRED.PATIENT DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION AT THE TIME OF CONTACT WITH DEXCOM TECHNICAL SUPPORT.