FDA Adverse Event Malfunction Summary report: N

HEARTWARE® VENTRICULAR ASSIST SYSTEM

MDR report key: 3984030 · Received August 6, 2014

Report

Report Number
3007042319-2014-00797
Event Type
Malfunction
Date Received
August 6, 2014
Date of Event
January 17, 2014
Report Date
July 8, 2014
Manufacturer
HEARTWARE
Product Code
DSQ
PMA / PMN Number
P100047
Removal / Correction Number
Z-1607-2014
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE BATTERIES WERE RETURNED; VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED IN ORDER TO EVALUATE THE PERFORMANCE OF THE BATTERY IN RELATION TO THE REPORTED EVENT. THOROUGH EXTERNAL VISUAL INSPECTION OF THE BATTERIES REVEALED NO SIGNS OF PHYSICAL DAMAGE OR CONTAMINATION. FUNCTIONAL TESTING REVEALED THAT (B)(4) CONTAINED A FAULTY CELL PAIR. AN INTERNAL INVESTIGATION (CAPA) HAS BEEN OPENED TO ADDRESS THE ISSUE.

Additional Manufacturer Narrative · 1

PRELIMINARY EVALUATION AND TESTING REVEALED THAT THE RETURNED BATTERY CONTAINED A FAULTY CELL. THIS FINDING WAS RE-ASSESSED PER OUR SOP REQUIREMENTS, AND DETERMINED TO BE REPORTABLE. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN THIRTY (30) DAYS OF COMPLETION OF THE INVESTIGATION. EVALUATION IN PROGRESS.

Description of Event or Problem · 1

APPROXIMATELY ELEVEN MONTHS POST HVAD IMPLANTATION, THE SITE REPORTED THAT THE PATIENT EXPERIENCED POWER SOURCE CHANGE IN THE CONTROLLER EARLIER THAN EXPECTED. THE BATTERIES WERE REMOVED FROM THE PATIENT AND A NEW SET WAS SUPPLIED. NO HARM OR INJURY TO THE PATIENT WAS REPORTED AS A RESULT OF THIS INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
459157 HEARTWARE® VENTRICULAR ASSIST SYSTEM CIRCULATORY ASSIST SYSTEM, BATTERY DSQ HEARTWARE

Patients

Seq Age Sex Outcome Treatment
1 (B)(4) - BATTERY