FDA Adverse Event Malfunction Summary report: N

1.55MM X 10MM FLUTED TWIST DRL

MDR report key: 3983972 · Received December 18, 2013

Report

Report Number
1045834-2013-15857
Event Type
Malfunction
Date Received
December 18, 2013
Report Date
August 9, 2012
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
K011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED BY DEPUY SYNTHES POWER TOOL. THE DEVICE WAS EVALUATED AND THE CUSTOMER'S COMPLAINT OF PACKAGING DEFECTS COULD NOT BE CONFIRMED. VISUAL INSPECTION ACCEPTANCE CRITERIA ARE "NO LOOSE FOREIGN MATERIAL (LFM) WHEN INSPECTED AT 1X MAGNIFICATION." NO LFM COULD BE FOUND ON DEVICE WHEN INSPECTED AT 10X MAGNIFICATION. ALSO, THE PACKAGE IS IN GOOD CONDITION WITH NO DEFECTS OF THE PEELABLE OR TERMINAL SEAL. IF ADD'L INFO BECOMES AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SENT. REF: (B)(4).

Description of Event or Problem · 1

REPORT RECEIVED FROM (B)(6) STATING THE DEVICE CONTAINED FOREIGN DEBRIS IN THE STERILE POUCH. THE DEVICE WAS NOT USED DURING SURGERY. THERE WAS NO PT/USER INJURY REPORTED. THE DATE OF EVENT IS UNK. THERE IS NO ADD'L INFO PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
663291 1.55MM X 10MM FLUTED TWIST DRL HBC DEPUY SYNTHES POWER TOOLS E493058930

Patients

Seq Age Sex Outcome Treatment
1