LIFEVEST WCD 4000 SYSTEM

MDR report key: 3983758 · Received June 25, 2014

Report

Report Number
3008642652-2014-01922
Event Type
Malfunction
Date Received
June 25, 2014
Date of Event
May 16, 2014
Report Date
June 24, 2014
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHARGER FAULT) HAS BEEN CONFIRMED. UPON INVESTIGATION THE BATTERY CHARGER/MODEM WAS UNABLE INVESTIGATION THE BATTERY CHARGER/MODEM WAS UNABLE TO CHARGE BATTERY PACKS. THE CAUSE FOR THE FAILURE WAS ISOLATED TO CONTAMINATION ON THE BATTERY PCA BOARD. THE ROOT CAUSE FOR THE CONTAMINATION WAS INGRESS OF AN UNK LIQUID. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY CHARGER/MODEM. THE PT RECEIVED A REPLACEMENT BATTERY CHARGER/MODEM.

Description of Event or Problem · 1

THE HUSBAND OF A (B)(6) FEMALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT SHE WAS RECEIVING BATTERY CHARGER FAULTS. THE PT WAS ISSUED A REPLACEMENT BATTERY CHARGER/MODEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
372190 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 62 YR