THERMOPHORE AUTOMATIC HEAT PACK
Report
- Report Number
- 1811605-2014-00151
- Event Type
- Malfunction
- Date Received
- June 25, 2014
- Date of Event
- June 2, 2014
- Report Date
- June 25, 2014
- Manufacturer
- BATTLE CREEK EQUIPMENT
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
OUR INSPECTION FOUND A SMALL CUT IN THE CORD NEAR THE SWITCH THAT COULD HAVE EXPOSED THE USER TO BARE WIRE. THE INSPECTION ALSO REVEALED CUSTOMER MISUSE AS OUR INSPECTION FOUND: FDA: CORD CUT/BURNED STRAIN RELIEF, SWITCH OK, PAD, BUNCHED, BENT/BROKEN LEAD, BENT/BROKEN THERMOSTAT, THERMOSTAT DISCOLORATION. THE PAD ALSO APPEARS TO HAVE BEEN LAID ON. THIS TELLS US THAT THE PAD WAS NOT BEING PROPERLY USED AS PER OUR INSTRUCTIONS IN THE MANUAL. CORD BREAKAGE USUALLY OCCURS WHEN THE CORD IS REPEATEDLY OVER BENT NEAR THE SWITCH. WE HAVE INITIATED A CAPA PROJECT ((B)(4)) TO REDUCE THE OCCURRENCE OF THIS ISSUE. A NEW JACKETED CORD DESIGN WAS LAUNCHED IN PRODUCTION ON (B)(4) 2014 THAT CLOSES (B)(4) PENDING A FUTURE EVALUATION OF THE EFFECTIVENESS OF THE SOLUTION.
CUSTOMER CALLED AND STATED THAT SPARKS CAME OUT OF THE CORD BY THE CASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 371576 | THERMOPHORE AUTOMATIC HEAT PACK | ELECTRIC MOIST HEAT PACK | IRT | BATTLE CREEK EQUIPMENT | 055 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |