FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3983694 · Received June 25, 2014

Report

Report Number
3008642652-2014-01924
Event Type
Malfunction
Date Received
June 25, 2014
Date of Event
May 20, 2014
Report Date
June 25, 2014
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (RESETTING) HAS BEEN CONFIRMED. UPON INVESTIGATION THE BATTERY CHARGER/MODEM WAS RESETTING. THE CAUSE FOR THE RESETS WAS ISOLATED TO A SHORTED Q1 CURRENT-CONTROLLING TRANSISTOR ON THE BEDSIDE PCA BOARD. THE ROOT CAUSE FOR THE SHORTED Q1 COMPONENT COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY CHARGER/MODEM. THE PATIENT RECEIVED A REPLACEMENT BATTERY/CHARGER/MODEM.

Description of Event or Problem · 1

THE DAUGHTER OF A (B)(6) FEMALE PATIENT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE PATIENT'S BATTERY CHARGER WAS RESETTING. THE PATIENT WAS ISSUED A REPLACEMENT BATTERY CHARGER/MODEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
371417 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION WCD 4000 NA