FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3983694
·
Received June 25, 2014
Report
- Report Number
- 3008642652-2014-01924
- Event Type
- Malfunction
- Date Received
- June 25, 2014
- Date of Event
- May 20, 2014
- Report Date
- June 25, 2014
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (RESETTING) HAS BEEN CONFIRMED. UPON INVESTIGATION THE BATTERY CHARGER/MODEM WAS RESETTING. THE CAUSE FOR THE RESETS WAS ISOLATED TO A SHORTED Q1 CURRENT-CONTROLLING TRANSISTOR ON THE BEDSIDE PCA BOARD. THE ROOT CAUSE FOR THE SHORTED Q1 COMPONENT COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY CHARGER/MODEM. THE PATIENT RECEIVED A REPLACEMENT BATTERY/CHARGER/MODEM.
Description of Event or Problem · 1
THE DAUGHTER OF A (B)(6) FEMALE PATIENT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE PATIENT'S BATTERY CHARGER WAS RESETTING. THE PATIENT WAS ISSUED A REPLACEMENT BATTERY CHARGER/MODEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 371417 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR |