ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3983691 · Received August 6, 2014

Report

Report Number
2531779-2014-22571
Event Type
Malfunction
Date Received
August 6, 2014
Report Date
July 31, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED, AND IT WAS EVALUATED BY PRODUCT ANALYSIS ON 09/11/2014 WITH THE FOLLOWING FINDINGS: VISUAL INSPECTION REVEALED THAT THE KEYPAD COVER WAS INTACT. EVALUATION REVEALED THAT THE UP AND DOWN ARROW KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE; THE REPORTED EVENT WAS DUPLICATED. THE KEYPAD COVER WAS REMOVED, AND EVIDENCE OF CONTAMINATION WAS FOUND UNDER ALL OF THE KEY CONTACTS. UNRELATED TO THE REPORTED EVENT, THE DISPLAY SCREEN WAS FOUND TO BE DIM AND DISCOLORED. ALSO UNRELATED TO THE REPORTED EVENT, THE CARTRIDGE COMPARTMENT WAS CRACKED BELOW THE GRIP PAD.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER ALLEGED THAT THE UP AND DOWN ARROW KEYPAD BUTTONS WERE UNDER RESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
461275 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 5 YR