FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3983678 · Received June 25, 2014

Report

Report Number
3008642652-2014-01925
Event Type
Malfunction
Date Received
June 25, 2014
Date of Event
May 21, 2014
Report Date
June 25, 2014
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WON'T POWER UP) HAS BEEN CONFIRMED. UPON INVESTIGATION THE BATTERY CHARGER/MODEM WAS UNABLE TO POWER UP. THE CAUSE FOR THE FAILURE WAS ISOLATED TO A DEFECTIVE POWER SUPPLY BRICK. THE ROOT CAUSE FOR THE DEFECTIVE POWER SUPPLY BRICK COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY CHARGER/MODEM. THE PATIENT RECEIVED A REPLACEMENT BATTERY CHARGER/MODEM.

Description of Event or Problem · 1

A ZOLL PATIENT SERVICE REPRESENTATIVE (PSR) CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT, WHILE FITTING A (B)(6) FEMALE PATIENT, THE PATIENT'S BATTERY CHARGER WAS NOT POWERING UP. THE PATIENT WAS ISSUED A REPLACEMENT BATTERY CHARGER/MODEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
371027 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 33 YR