FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3983625 · Received August 6, 2014

Report

Report Number
2531779-2014-22569
Event Type
Malfunction
Date Received
August 6, 2014
Report Date
July 26, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. DURING A VISUAL INSPECTION OF THE PUMP, MOISTURE WAS VISIBLE BEHIND THE PUMP DISPLAY LENS. THE BATTERY COMPARTMENT WAS ALSO OBSERVED TO BE CRACKED, AND THE BOLUS BUTTON COVER WAS TORN. NO EVIDENCE OF MOISTURE INGRESS WAS FOUND IN THE BATTERY COMPARTMENT. THE PUMP CASE WAS REMOVED AND CORROSION DUE TO MOISTURE WAS FOUND ON THE FORCE SENSOR PLATE. THE BOLUS BUTTON COVER WAS REMOVED AND CORROSION WAS FOUND ON THE BUTTON CONTACTS.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(6).

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (MOISTURE INGRESS) ISSUE. IT WAS REPORTED THAT EVIDENCE OF MOISTURE INGRESS WAS VISIBLE BEHIND THE PUMP DISPLAY LENS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
461115 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 14 YR