FDA Adverse Event Malfunction Summary report: N

V200 VENTILATOR

MDR report key: 3983606 · Received July 9, 2014

Report

Report Number
2031642-2014-00643
Event Type
Malfunction
Date Received
July 9, 2014
Report Date
June 16, 2014
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
CBK
PMA / PMN Number
K102054
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TH
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

THE INTERNATIONAL CUSTOMER REPORTED THE VENTILATOR CANNOT WORK ON AC POWER DURING NORMAL VENTILATION OPERATION. THE CUSTOMER REPORTED THE DEVICE WAS NOT IN USE ON A PT THEREFORE, THERE WAS NO PT INVOLVEMENT OR HARM. THE MFR FIELD SERVICE ENGINEER CONFIRMED THE REPORTED PROBLEM. THE MFR'S FIELD SERVICE ENGINEER EVALUATED THE DEVICE AND REPLACED THE POWER SUPPLY TO ADDRESS THE REPORTED PROBLEM. THE DEVICE PASSED ALL MFR REQUIRED TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
402659 V200 VENTILATOR VENTILATOR, CONTINUOUS CBK RESPIRONICS CALIFORNIA, INC. V200 NA

Patients

Seq Age Sex Outcome Treatment
1