FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 3983602 · Received June 25, 2014

Report

Report Number
9615739-2014-00250
Event Type
Malfunction
Date Received
June 25, 2014
Date of Event
May 28, 2014
Report Date
May 28, 2014
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE FOOT AND BRAKE CASTERS NOT HOLDING MOST LIKELY DUE TO NORMAL WEAR AND TEAR. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2013 AND 2014. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE FOOT AND BRAKE CASTERS TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECHNICIAN STATING THE FOOT AND BRAKES NOT HOLDING. THE BED WAS LOCATED AT THE ACCOUNT IN (B)(6). THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
371062 CAREASSIST BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM PLUVIGNER 1170

Patients

Seq Age Sex Outcome Treatment