FDA Adverse Event
Malfunction
Summary report: N
PACKER/CHANG IOL CUTTER
MDR report key: 3983601
·
Received June 25, 2014
Report
- Report Number
- 3019924-2014-00021
- Event Type
- Malfunction
- Date Received
- June 25, 2014
- Date of Event
- May 21, 2014
- Report Date
- June 24, 2014
- Manufacturer
- MICROSURGICAL TECHNOLOGY, INC.
- Product Code
- HNF
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE SCISSOR WAS RETURNED DIRTY AND CORRODED AT THE SITE OF THE BREAK INDICATING IT HAD NOT BE MAINTAINED ADEQUATELY PER THE INSTRUCTIONS FOR USE.
Description of Event or Problem · 1
THE FACILITY REPORTED THAT THE PACKER/CHANG IOL SCISSORS BROKEN WHEN THE SURGEON ATTEMPTED TO CUT A LENS. THE BROKEN PIECE OF THE SCISSOR FELL TO THE BACK OF THE EYE AND THE PATIENT WAS REFERRED TO A RETINAL SURGEON TO HAVE THE PIECE REMOVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 372240 | PACKER/CHANG IOL CUTTER | OPHTHALMIC SCISSOR | HNF | MICROSURGICAL TECHNOLOGY, INC. | DFH-0012 | 045234 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |