FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3983541 · Received August 6, 2014

Report

Report Number
2531779-2014-22528
Event Type
Malfunction
Date Received
August 6, 2014
Report Date
July 30, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1; DATE OF SUBMISSION: 10/07/2014 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED, AND IT WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE BLACK BOX DATA SHOWED NO EVIDENCE OF SHORT BATTERY LIFE OR EXCESSIVE BATTERY USAGE. THE PUMP POWERED ON WITH THE RETURNED BATTERY CAP, WHICH WAS ABLE TO SECURE TO THE PUMP PER THE INSTRUCTIONS FOR USE (IFU). VISUAL INSPECTION REVEALED THAT THE BATTERY COMPARTMENT WAS COMPLETELY INTACT. EVALUATION REVEALED THAT THE PUMP DREW ELECTRICAL CURRENT WITHIN THE SPECIFICATIONS. THE PUMP CASE WAS REMOVED, AND NO INTERNAL DEFECTS WERE FOUND. THE REPORTED EVENT WAS NOT DUPLICATED. DURING ANALYSIS, THE PUMP OPERATED ACCORDING TO THE SPECIFICATIONS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
459810 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 65 YR