FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3983536 · Received June 26, 2014

Report

Report Number
3008642652-2014-01965
Event Type
Malfunction
Date Received
June 26, 2014
Date of Event
May 18, 2014
Report Date
June 24, 2014
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEMS (DAMAGED ELECTRODE BELT) WAS CONFIRMED. UPON INVESTIGATION THE DISTRIBUTION NODE (DN) TO REAR THERAPY ELECTRODE (TE) CABLE WAS PULLED FROM THE STRAIN RELIEF. THIS RESULTED IN AN INTERMITTENT CONNECTION BETWEEN THE CABLE AND DN. THE ROOT CAUSE OF THE PULLED CABLES COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) FEMALE CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT HER LIFEVEST WAS BROKEN AND THE DEVICE INDICATED TO CALL ZOLL FOR A NEW ELECTRODE BELT. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
373963 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 52 YR