FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3983515
·
Received June 26, 2014
Report
- Report Number
- 3008642652-2014-01977
- Event Type
- Malfunction
- Date Received
- June 26, 2014
- Date of Event
- May 16, 2014
- Report Date
- June 25, 2014
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WON'T POWER UP) WAS CONFIRMED. AS RECEIVED, THE MONITOR WAS RESETTING AT START-UP. UPON INVESTIGATION, CRYSTAL OSCILLATOR Y500 WAS FOUND TO DEFECTIVE ON THE COMPUTER ANALOG BOARD. THE CAUSE OF THE RESETS WAS THE DEFECTIVE Y500. THE ROOT CAUSE OF THE DEFECTIVE Y500 WAS UNABLE TO BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE COMPONENT. THE PATIENT WAS ISSUED A REPLACEMENT MONITOR.
Description of Event or Problem · 1
A (B)(6) MALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT HIS MONITOR WOULD NOT POWER ON COMPLETED. THE PATIENT WAS ISSUED A REPLACEMENT MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 373961 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 33 YR |