FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3983515 · Received June 26, 2014

Report

Report Number
3008642652-2014-01977
Event Type
Malfunction
Date Received
June 26, 2014
Date of Event
May 16, 2014
Report Date
June 25, 2014
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WON'T POWER UP) WAS CONFIRMED. AS RECEIVED, THE MONITOR WAS RESETTING AT START-UP. UPON INVESTIGATION, CRYSTAL OSCILLATOR Y500 WAS FOUND TO DEFECTIVE ON THE COMPUTER ANALOG BOARD. THE CAUSE OF THE RESETS WAS THE DEFECTIVE Y500. THE ROOT CAUSE OF THE DEFECTIVE Y500 WAS UNABLE TO BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE COMPONENT. THE PATIENT WAS ISSUED A REPLACEMENT MONITOR.

Description of Event or Problem · 1

A (B)(6) MALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT HIS MONITOR WOULD NOT POWER ON COMPLETED. THE PATIENT WAS ISSUED A REPLACEMENT MONITOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
373961 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 33 YR