FDA Adverse Event Malfunction Summary report: N

LIGASURE ADVANCE PISTOL GRIP

MDR report key: 3983510 · Received June 26, 2014

Report

Report Number
1717344-2014-00557
Event Type
Malfunction
Date Received
June 26, 2014
Date of Event
April 30, 2014
Report Date
June 1, 2014
Manufacturer
COVIDIEN LP
Product Code
GEI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE CLEAR INSULATION WAS DAMAGED. THE DEVICE FAILED HIPOT TESTING. SEE SCANNED PAGE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE JAW WOULD NOT OPEN WHILE NOT ON PATIENT TISSUE. THERE WAS NO INJURY TO THE PATIENT. THE DEVICE WAS RETURNED TO COVIDIEN AND VISUAL INSPECTION DISCOVERED THAT THE CLEAR INSULATION WAS DAMAGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
373865 LIGASURE ADVANCE PISTOL GRIP LIGASURE ADVANCE PISTOL GRIP GEI COVIDIEN LP 32900139X

Patients

Seq Age Sex Outcome Treatment
1 UNK