TRACER HYBIRD WIRE GUIDE
Report
- Report Number
- 1037905-2014-00260
- Event Type
- Malfunction
- Date Received
- July 2, 2014
- Date of Event
- June 2, 2014
- Report Date
- June 2, 2014
- Manufacturer
- WILSON-COOK MEDICAL INC
- Product Code
- KOG
- PMA / PMN Number
- K910497
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
INVESTIGATION EVALUATION: OUR EVALUATION OF THE RETURNED DEVICE CONFIRMED THE REPORT. APPROXIMATELY 7 CM FROM THE DISTAL END IS AN APPROXIMATELY 7 CM SECTION OF EXPOSED BARE CORE WIRE. A 7 CM SECTION OF SEPARATED COATING WAS ALSO RETURNED WITH THE COMPLAINT. A NEW HYB-48015 WIRE GUIDE WAS ADVANCED INTO AN OLYMPUS JF130 3.2 ENDOSCOPE IN A SIMULATED UPPER GI POSITION WITH THE TIP IN A RETROFLEXED POSITION TO SIMULATE A WORST CASE SITUATION. THE TIP OF THE BALLOON WAS ADVANCED ONTO THE PREPOSITIONED NEW WIRE GUIDE WITH THE WIRE GUIDE EXITING THE BALLOON CATHETER AT THE IDE PORT. THE EXTRACTION BALLOON (FS-QEB-A) RETURNED WITH THE COMPLAINT DEVICE WAS ADVANCED OVER THE NEW WIRE GUIDE UNTIL IT EXITED THE DISTAL END OF THE SCOPE. THE BALLOON WAS THEN RETRACTED. THE NEW WIRE GUIDE WAS EXAMINED AND THERE WAS NO DAMAGE TO THE WIRE GUIDE. THE IDE PORT OF THE FS-QEB-A RETURNED WITH THE COMPLAINT DEVICE WAS MEASURED WITHIN THE SPECIFICATION. A PRODUCT-SPECIFIC DISCREPANCY THAT COULD HAVE CAUSED OR CONTRIBUTED TO THIS OBSERVATION WAS NOT OBSERVED DURING OUR LABORATORY ANALYSIS. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSIONS: A DEFINITIVE CAUSE FOR THIS OBSERVATION COULD NOT BE DETERMINED BECAUSE THE ACTUAL USE CONDITIONS COULD NOT BE DUPLICATED IN THE LABORATORY SETTING. DUE TO A VARIETY OF CLINICAL CONDITIONS SUCH AS PATIENT ANATOMY, ENDOSCOPE POSITION OR PROGRESSION OF DISEASE STATE, WE COULD NOT REPRODUCE THE ACTUAL CONDITIONS OF PRODUCT USAGE DURING OUR LABORATORY ANALYSIS. THIS LIMITS OUR ABILITY TO CONCLUSIVELY DETERMINE A CAUSE. THE INSTRUCTIONS FOR USE FOR THIS PRODUCT LINE INSTRUCT THE USER TO FLUSH THE WIRE GUIDE PRIOR TO REMOVAL FROM THE WIRE GUIDE HOLDER. THIS ACTIVITY WILL AID IN OPTIMAL PERFORMANCE OF THE WIRE GUIDE. FAILURE TO FLUSH THE WIRE GUIDE CAN RESULT IN DAMAGE TO THE WIRE GUIDE. THE INSTRUCTIONS FOR USE FOR THIS PRODUCT LINE INSTRUCT THE USER TO FLUSH ENDOSCOPE ACCESSORY CHANNEL AND/OR LUMEN OF DEVICE WITH STERILE WATER AND FOR BEST RESULTS, WIRE GUIDE SHOULD BE KEPT WET. THIS ACTIVITY WILL AID IN OPTIMAL PERFORMANCE OF THE WIRE GUIDE. FAILURE TO FLUSH THE ENDOSCOPE CHANNEL CAN RESULT IN DAMAGE TO THE WIRE GUIDE. THE INSTRUCTIONS FOR USE FOR THIS PRODUCT CAUTIONS THE USER THAT THIS PRODUCT IS COMPATIBLE WITH NON-METAL TIP DEVICES. USE OF THE WIRE GUIDE WITH METAL TIP DEVICES MAY COMPROMISE THE INTEGRITY OF THE EXTERNAL COATING ON THE WIRE GUIDE. THE REPORTED OBSERVATION CAN OCCUR IF THE WIRE GUIDE WAS USED WITH AN INCOMPATIBLE ACCESSORY DEVICE. IF ADDITIONAL PRESSURE IS APPLIED TO THE WIRE GUIDE AND/OR ACCESSORY DEVICE(S) WHILE MOVING THE WIRE GUIDE INSIDE THE ACCESSORY DEVICE(S), THIS COULD CONTRIBUTE TO WIRE GUIDE COATING DAMAGE. PRIOR TO DISTRIBUTION, ALL TRACER HYBRID WIRE GUIDES ARE SUBJECTED TO A VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK TRACER HYBRID WIRE GUIDE. AS THE COOK FUSION EXTRACTION BALLOON (FS-QEB-A) WAS ADVANCED OVER THE DISTAL END OF THE HYB-48015 WIRE GUIDE, THE COATING OF THE WIRE GUIDE STRIPPED OFF WHERE THE WIRE GUIDE WAS EXITING THE CATHETER, 9 CM FROM THE DISTAL TIP OF THE CATHETER. THE COATING WAS FOUND INSIDE THE CHANNEL OF THE ENDOSCOPE. THE PROCEDURE WAS COMPLETED WITH ANOTHER WIRE GUIDE. NO DEVICE OR DEVICE PORTION WAS LEFT INSIDE THE PATIENT. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 387546 | TRACER HYBIRD WIRE GUIDE | KOG, ENDOSCOPE AND/OR ACCESSORIES | KOG | WILSON-COOK MEDICAL INC | W3412842 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | REQUEST | OLYMPUS TJF-160VF ENDOSCOPE| COOK FUSION EXTRACTION BALLOON (FS-QEB-A)| COOK FUSION OMNI-TOME SPHINCTEROTOME |