FDA Adverse Event Malfunction Summary report: N

VASO VIEW HEMOPRO

MDR report key: 3983298 · Received July 3, 2014

Report

Report Number
2242352-2014-00740
Event Type
Malfunction
Date Received
July 3, 2014
Date of Event
June 6, 2014
Report Date
June 9, 2014
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
PMA / PMN Number
K052274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT YET BEEN RETURNED TO MAQUET CARDIAC SURGERY FOR EVALUATION. WE ARE FOLLOWING UP WITH THE CUSTOMER FOR THE RETURN OF THE DEVICE. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF THE DEVICE IS RECEIVED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE RECORDED IN THE LOT HISTORY. THERE ARE NO OTHER SIMILAR COMPLAINTS REPORTED AGAINST THIS BATCH. INTERNAL COMPLAINT NUMBER - (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE BTT BALLOON WOULD NOT HOLD AIR PRESSURE ON THE VASO HEMOPRO. IT APPEARED THAT THE VALVE WAS NOT WORKING PROPERLY. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
392001 VASO VIEW HEMOPRO ENDOSCOIC VESSEL HARVESTING GEI MAQUET CARDIOVASCULAR, LLC VH-3000 25096751

Patients

Seq Age Sex Outcome Treatment
1 54 YR