FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYS
MDR report key: 3983262
·
Received June 27, 2014
Report
- Report Number
- 3008642652-2014-01993
- Event Type
- Malfunction
- Date Received
- June 27, 2014
- Date of Event
- May 31, 2014
- Report Date
- June 26, 2014
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WIRES FROM BACK TE PAD NO LONGER CONNECTED) HAS BEEN CONFIRMED. AS RECEIVED, THE BELT FAILED INCOMING TESTING. UPON INVESTIGATION, THE DISTRIBUTION NODE (DN) TO REAR TE CABLE WAS PULLED FROM THE STRAIN RELIEF. THE CAUSE OF THE DAMAGED CABLE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE STRAINED CABLE. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
THE NURSE OF AN (B)(6) FEMALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THE WIRES FROM THE BACK THERAPY (TE) PAD WERE NO LONGER CONNECTED. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 376985 | LIFEVEST WCD 4000 SYS | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 86 YR |