FDA Adverse Event Malfunction Summary report: N

SINGLE ULTRASONIC DISSECTOR

MDR report key: 3983234 · Received June 27, 2014

Report

Report Number
1717344-2014-00560
Event Type
Malfunction
Date Received
June 27, 2014
Date of Event
May 30, 2014
Report Date
June 3, 2014
Manufacturer
COVIDIEN LP
Product Code
LFL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE INCIDENT ULTRASONIC DISSECTOR WAS RETURNED FOR EVALUATION. VISUAL INSPECTION OF THE DISPOSABLE HAND PIECE REVEALED THAT THE DEVICE HAD BEEN USED AND THE STATIC PART OF THE JAW HAD BROKEN OFF. THE BROKEN PIECE WAS RETURNED WITH THE REST OF DEVICE. THE REMAINING WAVE GUIDE AND THE BROKEN PIECE WERE INSPECTED UNDER MAGNIFICATION TO IDENTIFY THE POINT OF INITIAL CONTACT AND FRACTURE. IT WAS CONCLUDED THAT THE TITANIUM WAVEGUIDE FRACTURED DURING USE AND EVENTUALLY BROKE OFF. THE TITANIUM WAVEGUIDE PROBABLY CAME IN CONTACT WITH A HARD METALLIC OBJECT SUCH AS HEMOSTAT OR RETRACTOR AS EVIDENCED BY THE BREAK POINT AND METALLIC SCRAPING. THE USER'S GUIDE FOR THIS SYS WARNS: CONTACT BETWEEN THE ACTIVE BLADE AND OTHER METAL OBJECTS (HEMOSTATS, CLIPS, STAPLES, RETRACTORS, ETC.) MAY RESULT IN UNINTENDED DAMAGE TO TISSUE AND/OR DEVICE FAILURE.

Description of Event or Problem · 1

THE CUSTOMER ORIGINALLY REPORTED THAT THE DURING AN LADG PROCEDURE, A RED INDICATOR LIGHT WAS PROVIDED BY THE SYS AND THE DEVICE STOPPED WORKING. A NEW DISSECTOR WAS INSTALLED AND WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WAS NO PT INJURY. THE DEVICE WAS RETURNED FOR EVALUATION WITH A PIECE OF THE ACTIVE WAVE GUIDE DISENGAGED. THE PIECE WAS RETURNED WITH THE REMAINDER OF THE WAVEGUIDE. THE DEVICE MAY HAVE BROKEN AFTER THE PROCEDURE DURING CLEANING OR DURING TRANSPORT TO COVIDIEN BOULDER. ADDITIONAL QUESTIONS IN REGARD TO THE INCIDENT HAVE BEEN ASKED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
377632 SINGLE ULTRASONIC DISSECTOR ULTRASONIC DISSECTOR LFL COVIDIEN LP UNK

Patients

Seq Age Sex Outcome Treatment
1 UNK ULTRASONIC REUSABLE BATTERY PACK - SERIAL # UNK| ULTRASONIC REUSABLE GENERATOR-SERIAL #UNK