FDA Adverse Event Malfunction Summary report: N

GEMSTAR PAIN MGE SNG

MDR report key: 3983223 · Received July 3, 2014

Report

Report Number
9615050-2014-04304
Event Type
Malfunction
Date Received
July 3, 2014
Date of Event
June 6, 2014
Report Date
June 9, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K060806
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE SERVICE CTR WITH A REPORT FROM THE CUSTOMER CONTACT THAT DURING PREVENTATIVE MAINTENANCE TESTING THE DEVICE ALARMED WITH ERROR CODE 18/000/000 (HISTORY POINTER ERROR). THIS DID NOT INDICATE A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CTR, BATTERY LEAKAGE WAS NOTED IN THE BATTERY COMPARTMENT AND MIDDLE PRINTED CIRCUIT BOARD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
392358 GEMSTAR PAIN MGE SNG 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA