LIGASURE ADVANCE PISTOL GRIP
Report
- Report Number
- 1717344-2014-00559
- Event Type
- Malfunction
- Date Received
- June 27, 2014
- Date of Event
- April 30, 2014
- Report Date
- June 1, 2014
- Manufacturer
- COVIDIEN LP
- Product Code
- GEI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE CLEAR INSULATION WAS MISSING. THE DEVICE FAILED HIPOT TESTING. THE IFU STATES TO PREVENT DAMAGE TO THE FLEXIBLE INSULATION PROXIMAL TO THE JAWS, CONFIRM THE HANDLE IS FULLY CLOSED PRIOR TO INSERTION INTO AND EXTRACTION FROM THE CANNULA. USE THE APPROPRIATELY SIZED CANNULA TO ALLOW FOR EASY INSERTION AND EXTRACTION OF THE INSTRUMENT. FAILURE TO DO SO MAY IMPACT THE INTEGRITY OF THE FLEXIBLE INSULATION. CANNULAS WITH HARD, NON-BEVELED OPENINGS MAY CAUSE THE FLEXIBLE INSULATION TO RETRACT, WHICH MAY COMPROMISE THE INSULATION. IF RETRACTION OCCURS, THE INSTRUMENT MUST BE DISCARDED. DO NOT ATTEMPT TO CLEAN THE FLEXIBLE INSULATION. CLEANING MAY DAMAGE INSULATION. THE JAWS OF THE INCIDENT DEVICE WERE STUCK SHUT. THE KNIFE IS TRAPPED AND CONTAINED WITHIN THE JAWS, WHICH KEPT THE JAWS FROM OPENING. THIS CAN OCCUR WHEN THE BLADE IS EXTENDED AND THE JAWS ARE NOT COMPLETELY CLOSED. THIS ALLOWS THE KNIFE TRACK TO OPEN TOO WIDE AND THE BLADE MOVES OUT OF ITS TRACK. THE IFU CAUTIONS TO CONFIRM THAT THE JAWS HAVE REACHED THE CLOSED POSITION AND ARE LOCKED BEFORE ACTIVATING THE CUTTER.
THE CUSTOMER REPORTED THAT THE JAW WOULD NOT OPEN WHILE NOT ON PT TISSUE DURING A LAPAROSCOPIC COLORECTAL PROCEDURE. THERE WAS NO INJURY TO THE PT. THE DEVICE WAS RETURNED TO COVIDIEN AND VISUAL INSPECTION DISCOVERED THAT THE CLEAR INSULATION WAS MISSING. THE CUSTOMER REPORTED THAT NOTHING FELL INTO THE PT'S CAVITY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 378407 | LIGASURE ADVANCE PISTOL GRIP | LIGASURE VESSEL SEALING SYS | GEI | COVIDIEN LP | 32560008X |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |