FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3983194 · Received July 3, 2014

Report

Report Number
8020893-2014-01571
Event Type
Malfunction
Date Received
July 3, 2014
Date of Event
January 1, 2014
Report Date
June 5, 2014
Manufacturer
COVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) EVALUATED THE DEVICE, AND VERIFIED THE MALFUNCTION. THE CSE REPLACED THE TOUCHFRAME CABLE ASSEMBLY AND GRAPHIC USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB). THE UNIT PASSED ALL TESTS AND CALIBRATIONS, AND IT WAS OBSERVED TO BE OPERATING WITHIN MFG SPECS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED BY A CUSTOMER IN THE USA THAT AN 840 VENTILATOR GRAPHICAL USER INTERFACE (GUI) HAD AN INOPERABLE DISPLAY. THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
392407 840 VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR 840

Patients

Seq Age Sex Outcome Treatment
1