FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3983187
·
Received July 3, 2014
Report
- Report Number
- 8020893-2014-01588
- Event Type
- Malfunction
- Date Received
- July 3, 2014
- Date of Event
- June 5, 2014
- Report Date
- June 6, 2014
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) EVALUATED THE DEVICE, AND VERIFIED THE MALFUNCTION. THE CSE REPLACED THE GRAPHIC USER INTERFACE (GUI), CENTRAL PROCESSING UNIT (CPU), PRINTED CIRCUIT BOARD (PCB), BACKLIGHT INVERTER, PRINTED CIRCUIT BOARD (PCBS), AND UPGRADED THE SOFTWARE TO THE CURRENT REVISION. THE UNIT PASSED ALL TESTS AND CALIBRATIONS, AND IT WAS OBSERVED TO BE OPERATING WITHIN MFG SPECS. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED BY A CUSTOMER IN THE (B)(6) THAT AN 840 VENTILATOR GRAPHIC USER INTERFACE (GUI) BOTTOM DISPLAY WAS SCRAMBLED. THERE WAS NO PT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 392177 | 840 VENTILATOR | CONTINUOUS VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |