FDA Adverse Event Malfunction Summary report: N

TPS HANDPIECE CORD

MDR report key: 3983068 · Received August 6, 2014

Report

Report Number
0001811755-2014-02807
Event Type
Malfunction
Date Received
August 6, 2014
Date of Event
July 21, 2014
Report Date
July 21, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
EIA
PMA / PMN Number
K032117
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING THE DEVICE EVALUATION, THE TECHNICIAN DETERMINED THE CABLE¿S INTERNAL WIRING WAS DAMAGED. THE DEVICE WAS DISCARDED BY THE MANUFACTURER.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY THE TPS HANDPIECE CORD CAUSED A BIAS CURRENT MESSAGE TO BE DISPLAYED ON THE CONSOLE, SIGNALING A CONDITION OCCURRED IN WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. IT WAS REPORTED THAT THERE WAS NO ASSOCIATED PROCEDURE. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
461230 TPS HANDPIECE CORD UNIT, OPERATIVE DENTAL EIA STRYKER INSTRUMENTS-KALAMAZOO 13089

Patients

Seq Age Sex Outcome Treatment
1