FDA Adverse Event Injury Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3983047 · Received August 6, 2014

Report

Report Number
2531779-2014-22506
Event Type
Injury
Date Received
August 6, 2014
Date of Event
July 10, 2014
Report Date
July 13, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/05/2014 WITH THE FOLLOWING FINDINGS: THE PUMP BLACK BOX WAS REVIEWED AND FOUND EVIDENCE OF UNEXPLAINED POWER EVENTS OCCURRING ON (B)(6) 2014 AT 17:27; THE PUMP BLACK BOX SHOWED THAT WHEN THE PUMP WAS POWERED BACK ON, THE TIME AND DATE WAS MANUALLY SET INCORRECTLY TO (B)(6) 2014 AT 03:28. THE PUMP INSULIN DELIVERY WAS RESUMED AT 07:36. A REVIEW OF THE TOTAL DAILY DOSE HISTORY SHOWED THAT THE DAILY INSULIN DELIVERY TOTALS CORRECTLY REFLECTED THE USER PROGRAMMED BASAL RATES. NO ALARMS OR PUMP CONDITIONS INDICATING A MALFUNCTION WERE RECORDED IN BLACK BOX OR ALARM HISTORY. A 29 HOUR FLOW ACCURACY TEST WAS PERFORMED AND THE PUMP WAS FOUND TO BE DELIVERING WITHIN SPECIFICATIONS. NO DELIVERY RELATED DEFECTS WERE FOUND DURING TESTING.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS ALLEGING THAT THE PATIENT¿S BLOOD GLUCOSE DROPPED TO 29 MG/DL REQUIRING A NURSE TO ¿REVIVE¿ THE PATIENT FROM PASSING OUT. THE REPORTER INDICATED THAT THEY DID NOT REALIZE THAT THE PUMP WAS ¿BROKEN¿. THE PUMP WAS UNABLE TO BE REVIEWED AT THE TIME OF THE CALL. THIS REPORT IS MADE BASED ON THE ALLEGATION THAT THE PATIENT EXPERIENCED HYPOGLYCEMIA ASSOCIATED WITH AN ALLEGATION THAT THE PUMP WAS MALFUNCTIONING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
461113 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 9 YR Life Threatening| R