ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-22506
- Event Type
- Injury
- Date Received
- August 6, 2014
- Date of Event
- July 10, 2014
- Report Date
- July 13, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PATIENT
Narratives
THE PRODUCT HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/05/2014 WITH THE FOLLOWING FINDINGS: THE PUMP BLACK BOX WAS REVIEWED AND FOUND EVIDENCE OF UNEXPLAINED POWER EVENTS OCCURRING ON (B)(6) 2014 AT 17:27; THE PUMP BLACK BOX SHOWED THAT WHEN THE PUMP WAS POWERED BACK ON, THE TIME AND DATE WAS MANUALLY SET INCORRECTLY TO (B)(6) 2014 AT 03:28. THE PUMP INSULIN DELIVERY WAS RESUMED AT 07:36. A REVIEW OF THE TOTAL DAILY DOSE HISTORY SHOWED THAT THE DAILY INSULIN DELIVERY TOTALS CORRECTLY REFLECTED THE USER PROGRAMMED BASAL RATES. NO ALARMS OR PUMP CONDITIONS INDICATING A MALFUNCTION WERE RECORDED IN BLACK BOX OR ALARM HISTORY. A 29 HOUR FLOW ACCURACY TEST WAS PERFORMED AND THE PUMP WAS FOUND TO BE DELIVERING WITHIN SPECIFICATIONS. NO DELIVERY RELATED DEFECTS WERE FOUND DURING TESTING.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS ALLEGING THAT THE PATIENT¿S BLOOD GLUCOSE DROPPED TO 29 MG/DL REQUIRING A NURSE TO ¿REVIVE¿ THE PATIENT FROM PASSING OUT. THE REPORTER INDICATED THAT THEY DID NOT REALIZE THAT THE PUMP WAS ¿BROKEN¿. THE PUMP WAS UNABLE TO BE REVIEWED AT THE TIME OF THE CALL. THIS REPORT IS MADE BASED ON THE ALLEGATION THAT THE PATIENT EXPERIENCED HYPOGLYCEMIA ASSOCIATED WITH AN ALLEGATION THAT THE PUMP WAS MALFUNCTIONING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 461113 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 9 YR | Life Threatening| R |