FDA Adverse Event Malfunction Summary report: N

LOANER SYSTEM 6 DT ROTARY

MDR report key: 3983035 · Received August 6, 2014

Report

Report Number
0001811755-2014-02804
Event Type
Malfunction
Date Received
August 6, 2014
Date of Event
July 18, 2014
Report Date
July 18, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED RUN ON WAS CONFIRMED BY A MANUFACTURER REPAIR TECHNICIAN THROUGH FUNCTIONAL EVALUATION. UPON DISASSEMBLY, IT WAS FOUND THAT THE EBOX WAS IN NEED OF REPLACEMENT, WHICH CAN CAUSE THE REPORTED EVENT.

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY THE DEVICE RAN WITHOUT USER ACTIVATION AFTER THE TRIGGER WAS RELEASED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY, THE DEVICE RAN WITHOUT USER ACTIVATION AFTER THE TRIGGER WAS RELEASED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.