ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-22501
- Event Type
- Injury
- Date Received
- August 6, 2014
- Report Date
- July 24, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PATIENT
Narratives
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/27/2014 WITH THE FOLLOWING FINDINGS: THERE WAS ONE OCCLUSION ALARM OBSERVED IN THE BLACK BOX; THE FORCE AT TIME OF OCCLUSION IS HIGH. DAILY INSULIN DELIVERY TOTALS APPEAR INCONSISTENT DUE TO TIME/DATE ISSUE. THE REWIND, LOAD CARTRIDGE, AND PRIME STEPS PERFORMED SUCCESSFULLY WITH NO ALARMS. FORCE SENSOR CALIBRATION TEST CONFIRMED SENSOR IS DETECTING CORRECT FORCE AT 5LBS. THE PUMP EXERCISED FOR 24 HOURS WITH NO OCCLUSIONS OCCURRING. PUMP PASSED DELIVERY ACCURACY TEST AND FOUND TO BE DELIVERING WITHIN REQUIRED SPECIFICATIONS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE REPORTER CONTACTED ANIMAS ON (B)(6) 2014 ALLEGING AN OCCLUSION ISSUE. THE REPORTER STATED THAT THE PATIENT HAD A BLOOD GLUCOSE (BG) OF 573MG/DL WITH ABDOMINAL PAIN/VOMITING AND WAS TAKEN TO THE EMERGENCY ROOM AND DIAGNOSED WITH DIABETIC KETOACIDOSIS. THE PATIENT WAS TREATED WITH INSULIN VIA INJECTION AND IV FLUIDS. THIS REPORT IS BEING MADE DUE TO THE BG EXCURSION THE PATIENT EXPERIENCED DUE TO AN ALLEGED OCCLUSION ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 461053 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR | Hospitalization |