FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 3983003 · Received August 6, 2014

Report

Report Number
0001811755-2014-02796
Event Type
Malfunction
Date Received
August 6, 2014
Date of Event
July 14, 2014
Report Date
July 14, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
ERL
PMA / PMN Number
K112593
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN WAS ABLE TO DUPLICATE THE REPORTED EVENT OF HEATING UP AND NOTED THAT THE COUPLER WAS WORN.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY THE CORE IMPACTION DRILL WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
461508 CORE IMPACTION DRILL DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE ERL STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1