FDA Adverse Event Malfunction Summary report: N

COYOTE? ES

MDR report key: 3982790 · Received August 6, 2014

Report

Report Number
2134265-2014-04614
Event Type
Malfunction
Date Received
August 6, 2014
Date of Event
July 14, 2014
Report Date
July 14, 2014
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LIT
PMA / PMN Number
K080982
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR.; VISUAL INSPECTION SHOWED THE HUB OF THE DEVICE WAS PROTRUDING OUT THE SIDE OF THE OPENED PRODUCT POUCH 4MM. THE CARRIER TUBE (HOOP) WAS POSITIONED IN THE PRODUCT POUCH WITH THE HUB OF THE DEVICE UNDER THE TYVEK SECTION AGAINST THE EDGE OF THE VENDOR SEAL. INSPECTION OF THE REMAINDER OF THE DEVICE AND PACKAGING REVEALED NO DAMAGE OR IRREGULARITIES. NO ISSUES WERE IDENTIFIED DURING THE PRODUCT ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS CONSIDERED HANDLING DAMAGE AS THE EVENT OCCURRED PRIOR TO PATIENT CONTACT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING UNPACKING DEVICE CONTAMINATION OCCURRED. A 4MM X 40MM X 145CM COYOTE¿ ES BALLOON CATHETER WAS REMOVED FROM THE PACKAGE WHEN THE DEVICE FLIP TO THE SIDE AND WAS UNSTERILE. THE DEVICE WAS NOT USED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING UNPACKING DEVICE CONTAMINATION OCCURRED. A 4MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS REMOVED FROM THE PACKAGE WHEN THE DEVICE FLIP TO THE SIDE AND WAS UNSTERILE. THE DEVICE WAS NOT USED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
459136 COYOTE? ES CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL LIT BOSTON SCIENTIFIC - MAPLE GROVE H74939134404010 16920616

Patients

Seq Age Sex Outcome Treatment
1