FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3982767 · Received July 24, 2014

Report

Report Number
1720753-2014-06362
Event Type
Malfunction
Date Received
July 24, 2014
Date of Event
July 11, 2014
Report Date
July 25, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT SCAN. THIS RESULTED A TOTAL LOSS OF IMAGING FUNCTIONALITY. THIS COULD RESULT IN THE DELAY OR CANCELLATION OF A PROCEDURE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
434889 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1