FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3982766 · Received July 24, 2014

Report

Report Number
1720753-2014-06364
Event Type
Malfunction
Date Received
July 24, 2014
Date of Event
July 11, 2014
Report Date
July 25, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE FUSES AND CONNECTORS WERE RESEATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE MONITOR HAD NO SIGNAL. THE FE REPORTED IT WAS A NO BOOT ISSUE. THIS RESULTED IN A TOTAL LOSS OF IMAGING FUNCTIONALITY. THIS COULD CAUSE THE DELAY OR CANCELLATION OF A PROCEDURE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.