FDA Adverse Event Malfunction Summary report: N

PLMA DVC V11.51 1 N

MDR report key: 3982717 · Received July 24, 2014

Report

Report Number
9615050-2014-04572
Event Type
Malfunction
Date Received
July 24, 2014
Date of Event
June 30, 2014
Report Date
June 30, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K070398
Removal / Correction Number
FA210-07
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

DURING TESTING, THE DEVICE ALARMED WITH AN E321 (BATTERY CHARGER TIMEOUT) ERROR CODE. THE CUSTOMER CONTACT'S REPORT THAT THE DEVICE ALARMED FOR WARNING REPLACE BATTERY WAS NOTED IN THE DEVICE HISTORY, BUT WAS NOT DUPLICATED DURING TESTING. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT FROM THE CUSTOMER CONTACT THAT THE DEVICE ALARMED WITH WARNING REPLACE BATTERY. NO ADD'L INFO WAS PROVIDED. THIS IS NOT A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER, THE DEVICE ALARMED WITH AN E321 (BATTERY CHARGER TIMEOUT) ERROR CODE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
435109 PLMA DVC V11.51 1 N 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA PLUM A+ SOFTWARE MODULE: LIST #12097,| SN (B)(4)