FDA Adverse Event Malfunction Summary report: N

3.5MM TI CORTEX SCREW 32MM

MDR report key: 3982694 · Received August 6, 2014

Report

Report Number
3009450884-2014-10047
Event Type
Malfunction
Date Received
August 6, 2014
Date of Event
July 7, 2014
Report Date
July 7, 2014
Manufacturer
SYNTHES GRENCHEN
Product Code
HWC
PMA / PMN Number
PK131186
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. ADDITIONAL PRODUCT CODE: HRS. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: A MANUFACTURING INVESTIGATION WAS CONDUCTED. THE REPORT INDICATES THAT THE CAUSE OF THE FAILURE IS NOT DUE TO ANY MANUFACTURING NON-CONFORMANCES, THE REASON FOR A BREAKAGE IS THE INSERTION WITHOUT PRE DRILLING AND/OR TOO MUCH APPLIED MECHANICAL FORCE HAS RESULTED IN THE BREAKAGE. NO PRODUCT FAULT COULD BE DETECTED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: THE SCREW WAS INSERTED WITHOUT PREDRILLING AND BROKE AT THE SCREW HEAD. THIS HAPPENED DURING OPERATION, NO RESIDUE IN PATIENT AND THE BROKEN SCREW WAS REMOVED WITH SCREW EXTRACTION SET. THIS COMPLAINT INVOLVES ONE DEVICE. THIS REPORT IS 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
459400 3.5MM TI CORTEX SCREW 32MM SCREW,FIXATION,BONE HWC SYNTHES GRENCHEN 8961419

Patients

Seq Age Sex Outcome Treatment
1 14 YR