OT VERIO IQ METER
Report
- Report Number
- 2939301-2014-19823
- Event Type
- Malfunction
- Date Received
- August 6, 2014
- Report Date
- July 29, 2014
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 ¿ (10/09/2014).THE PATIENT¿S TEST STRIPS HAVE BEEN RETURNED ON 9/5/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 9/29/2014 WITH THE FOLLOWING FINDINGS: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT FAILED TESTING. WHEN TEST STRIPS WERE TESTED WITH CONTROL SOLUTION, AN ERROR 4 WAS OBSERVED WITH NO ASSIGNABLE CAUSE FOUND. IN ADDITION, A SECONDARY ISSUE WAS NOTED WHEN THE TEST STRIPS WERE FOUND TO HAVE RESULTS GREATER THAN +50% OF THE CONTROL SOLUTION WHEN TESTED WITH CONTROL SOLUTION. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2014, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING ERROR 4. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.